SAV-DO700 Optical dissolved oxygen meter

SAV-DO700 Optical dissolved oxygen meter


SAV-DO700 dissolved Oxygen meter adopts fluorescence method to measure dissolved oxygen.The cap of the sensor is coated with a luminescent material. Blue light from an LED illuminates the luminescent chemical. The luminescent chemical instantly becomes excited and releases red light. The time and intensity of red light are inversely proportional to the concentration of oxygen molecules, So the concentration of oxygen molecules is calculated.



Features

Range: 0-20mg/L,0-200%,0-400hPa

Resolution: 0.01mg/L,0.1%,1hPa

Output signal: 4~20mA; Relay; RS485

Power supply: AC220V±10%; 50Hz/60Hz

SPECIFICATIONS

Product Dissolved oxygen meter
Model SAV-DO700
Measure range 0-20mg/L, 0-20ppm, 0-45deg C
Accuracy Resolution:±3%, Temperature: ±0.5℃
Pressure range ≤0.3Mpa
Calibration Automatic air calibration,Sample calibration
Sensor material SUS316L+PVC (Ordinary Version),
Titanium Alloy (Seawater Version)
O-ring: Fluoro-rubber; Cable: PVC
Cable length Standard 10-Meter Cable, Max : 100m
Display 128 * 64 dot matrix LCD with LED backlight
Output 4-20mA(Max three-way);
RS485 MODBUS;
Rlay output(Max three-way);
Power supply AC220V, 50Hz,(optional 24V)


Description


SAV-DO700 Dissolved oxygen meter measures the dissolved oxygen by the fluorescence method, and the emitted blue light is irradiated on the phosphor layer. The fluorescent substance is stimulated to emit red light, and the oxygen concentration is inversely proportional to the time when the fluorescent substance returns to the ground state. By using this method to measure the dissolved oxygen, it will not produce oxygen consumption, thus assuring data stability, reliable performance, no interference, and simple installation and calibration.


PRODUCT ADVANTAGES

· The sensor uses a new oxygen-sensitive membrane type with NTC temperature compensation, resulting in improved stability and repeatability of measurement results.

· When measuring, there is no need for flow rate or stirring, and it will not produce oxygen consumption.

· Our fluorescence technology is a breakthrough that requires no membrane or electrolyte and requires very minimal maintenance.

· There is a self-diagnosis feature included to guarantee the precision of the information.

· The factory calibration is sufficient for a year and field calibration can also be carried out if necessary.

·This device is equipped with a digital sensor that has high anti-jamming capabilities and can transmit data over a far distance.

· The digital signal output is standard and allows integration and networking with other equipment without the need for a controller.

· This sensor is plug-and-play and can be installed quickly and easily.

· Keep industrial controlled doors closed to prevent equipment from halting.